5. Diagnostics & Maintenance

5. Diagnostics & Maintenance

Output does not switch

  • Confirm that SmartSampler2 and the PowerExpander are powered and connected.
  • Check the output assignment and test the selected port in SmartConfig2.
  • Disconnect power, then inspect the port fuse and replace it only with a suitable rating no greater than 7.5 A.
  • Inspect the load and cable for a short circuit, overload, loose connection, or reversed polarity before restoring power.

Routine maintenance

  • Inspect connectors, seals, cables, and the enclosure for damage or contamination.
  • Keep unused ports capped.
  • Clean the exterior with a damp cloth; do not use solvents or high-pressure cleaning.
  • Record repeated fuse failures and the connected load details before requesting support.