To verify that time updates via a connected transmitter are working correctly, follow these steps:
- Set a Test Time:
Manually set the SmartSampler to a “wrong” time to ensure you can observe the update. - Adjust the Offset:
- Navigate to Basic Settings and temporarily change the applied time offset.
- Apply the new offset.
- Synchronice the Time so the offest will be applied
- Revert the Offset:
- Restore the original time offset.
- Power Cycle the Device:
- Disconnect the main power (USB can remain connected).
- Wait for at least 3 seconds.
- Reconnect the power.
- Verify the Update:
- Use the Serial Monitor –> Cellular to observe the time sync. You should see messages like after it connected to the network:
- 2025/05/08 02:56:12, update RTC
- Set time to: 25/05/08,10:56:13+08 ignoring the offset from network
- 2025/05/08 08:56:13, Time was off for more than 10 seconds, it has been set from cellular to: 2025/05/08 08:56:13
- 2025/05/08 08:56:13, time was off for more than 10 minutes (this would restart the system if it is running already)
- Use the Serial Monitor –> Cellular to observe the time sync. You should see messages like after it connected to the network:
- Handling Network Limitations:
- If the transmitter cannot obtain the time from the network, you may see:
- 2025/05/08 08:27:38, update RTC
- Cellular time is not set, skipping setting time
- If the transmitter cannot obtain the time from the network, you may see:
Note: Not all network providers support automatic time updates. If this message appears consistently, it might indicate that the network lacks this feature.