6.1.3 FlowControl

Steps to Calibrate:

1. Reset Calibration

If your current calibration is significantly inaccurate, you can start fresh:

  • Click Reset to remove all previously stored calibration data and return to default settings.

2. Initial Test Run (after reset; units typically reflect milliliters)

  • Enter a Target value (e.g., 25 units).
  • Set the Real value to the same number (e.g., 25).
  • Click Start to begin the fill process.
  • The system will dispense the selected volume into your measurement container.
  • Observe and measure the actual output (e.g., using a graduated cylinder).

3. Enter Real Output and Apply Correction

  • Measure the actual volume dispensed (e.g., 15 units).
  • Enter this measured value into the Real field.
  • Click Apply to calculate and store the correction factor based on the difference between Target and Real.

The calibration adjusts internal flow-to-volume conversion to better match your system’s actual performance.